皮肤温度控制模块

#moorVMS-HEAT

Skin heater module offers flexible and reproducible skin heating protocols for flow/ oxygen response assessments

  • We can't recommend Moor instruments highly enough. The technology is at the cutting edge and the support second to none.

    Paul Sumners, PhD
    London South Bank University

  • I cannot rate the company or the staff highly enough.

    Jim House, PhD
    University of Portsmouth

  • We have found Moor equipment to be extremely dependable and innovative.

    Dean L. Kellogg, Jr., MD, Ph.D
    University of Texas Health Science Center

  • Laser Doppler Imager is a standard accurate method we now use in our cerebral blood flow and brain perfusion in our laboratory.

    Momoh A. Yakubu, PhD
    Texas Southern University

  • Moor Instruments have consistently provided excellent help and support for my research.

    Kim Gooding, PhD
    University of Exeter Medical School

  • In a nutshell, moorFLPI-2 is the most user-friendly system for studying cerebral blood flow regulation in rodents.

    Chia-Yi (Alex) Kuan, MD, PhD
    Emory University School of Medicine

  • I expect to be using Moor Instrument’s technology for many years to come!

    Faisel Khan, PhD
    Ninewells Hospital & Medical School

  • It goes without saying that the company's imaging technology itself is superb!

    Gourav Banerjee
    Leeds Beckett University

moorVMS-HEAT Skin Heater Controller: Advanced Control for Research and Clinical Studies

Moor Instruments’ moorVMS-HEAT Skin Heater Controller sets the benchmark in skin heating accuracy for Chinese research labs, universities, and clinical research studies.

Specially engineered to work effectively with our laser Doppler blood flow monitors and the moorVMS-OXY tissue oxygenation monitors, it provides reliable, reproducible skin tissue heating up to 45ºC, enhancing the precision and quality of your research.

Why Researchers Choose moorVMS-HEAT:

  • Simultaneous dual-channel independent heating capability.
  • User-friendly standalone or computer-managed protocols.
  • Automated analysis and data management.
  • Reliable warranty support, three years standard (extendable).
  • Robust connectivity via USB and analogue outputs.

Request a Call Back | Download Brochure | Call Moor Instruments today

以下产品可在线购买并可用于moorVMS-HEAT


本节列出了对客户moorVMS-HEAT常见问题的解答。 如果您的问题不在其中,请将问题电邮给我们。 我们将乐于帮助!


Q. What is the moorVMS-HEAT used for?
A. The moorVMS-HEAT can be used to regulate the blood flow measurement temperature or it can be used to affect reproducible skin temperature changes as a stimulus for blood flow change.

Q. Is the moorVMS-HEAT a stand-alone skin-heating controller?
A. Yes, the moorVMS-HEAT can be used in a stand-alone mode or it can be controlled by moorVMS-PC software. This allows convenient connection of a number of skin heating and iontophoresis modules and full flexibility in the protocol set up.

Q. Can the moorVMS-HEAT be remotely controlled?
A. Yes, the moorVMS-HEAT can be remotely controlled by the moorVMS-PC software, where individual protocols detailing temperature and duration can be programmed.

Q. What is the temperature range of the moorVMS-HEAT?
A. The moorVMS-HEAT has a measurement range of 5-50°C and a heating range of 20-45°C (or ambient to 45°C, if the ambient temperature is greater than 20°C). Important: Ensure the temperature selected is appropriate for the subject to avoid risk of local burns.

Q. Can both channels of the moorVMS-HEAT be set to different temperatures - i.e. controlled independently?
A. Yes, the moorVMS-HEAT probes can be at different temperatures within each time period.

Q. Do you have to have two probes connected to the moorVMS-HEAT for it to work?
A. No, the moorVMS-HEAT will work with only one probe connected.

Q. Do I need to calibrate the moorVMS-HEAT probes?
A. There is no user calibration of the heater probes and new probes are supplied calibrated. We recommend annual factory servicing in order to ensure the system is operating within specification.

Q. When buying new probes for my moorVMS-HEAT, do I have to return the moorVMS-HEAT to calibrate the new probes?
A. No, new probes are calibrated by Moor Instruments without the need of returning the moorVMS-HEAT.

Q. Can I iontophorese and heat at the same time?
A. Yes, use the moorVMS-HEAT VHP2 digit-heating probe in the ION1r Iontophoresis chamber. The moorVMS-PC software allows simultaneous control of both skin heating and iontophoresis protocols.

Q. What optic probes do I use with the moorVMS-HEAT?
A. When using a moorVMS-LDF, you need to use a VP12. For the DRT4 you need to use a DP12-V2 and when using a moorLAB you need to use a MP12-V2.

Q. What adhesive discs do I need to order for my moorVMS-HEAT probes?
A. VHP1 you need to order SHAD.
VHP2 you need to order PAD.
VHP3 you need to order IAD or VHP-FP.

Q. Does the moorVMS-HEAT come with a warranty?
A. Yes, there is a 3-year warranty included with the purchase of a moorVMS-HEAT (the heating probes are covered by a 6 months warranty).

There are numerous references where our skin heaters are cited. The list below is a small selection. Please contact us for reference lists on your chosen subject.


Brunt, V. E., Miner, J. A., Meendering, J. R., Kaplan, P. F., and Minson, C. T., (2012).
17-β estradiol and progesterone independently augment cutaneous thermal hyperemia but not reactive hyperemia.
Microcirculation. 2011 Jul; 18(5): 347–355.
Weblink

Brunt, V. E., Eymann, T. M., Francisco, M. A., Howard, M. J., and Minson, C. T., (2016).
Passive heat therapy improves cutaneous microvascular function in sedentary humans via improved nitric oxide-dependent dilation.
Journal of Applied Physiology Published 14 July 2016 Vol. no. , DOI: 10.1152/japplphysiol.00424.2016.
Weblink

Van Duijnhoven, N. T. L., Janssen, T. W. J., Green, D. J., Minson, C. T., Hopman, M. T. E., and Thijssen, D. H. J., (2009).
Effect of functional electrostimulation on impaired skin vasodilator responses to local heating in spinal cord injury.
Journal of applied physiology (Bethesda, Md.: 1985), 106(4), pp.1065–71.
Weblink

Dupont, J. J., Farquhar, W. B., Townsend, R. R., and Edwards, D. G., (2011).
Ascorbic acid or L-arginine improves cutaneous microvascular function in chronic kidney disease.
Journal of applied physiology (Bethesda, Md.: 1985), 111(6), pp.1561–7.
Weblink

Harvey, J. C., Roseguini, B. T., Goerger, B., Fallon, E. A., and Wong, B. J., (2016).
Acute thermotherapy prevents impairments in cutaneous microvascular function induced by a high fat meal.
Journal of Diabetes Research. Submission Accepted 5th July 2016.
Weblink

Heimhalt-El Hamriti, M., Schreiver, C., Noerenberg, A., Scheffler, J., Jacoby, U., Haffner, D., and Fischer, D.C., (2013).
Impaired skin microcirculation in paediatric patients with type 1 diabetes mellitus.
Cardiovascular diabetology, 12(1), p.115.
Weblink

Holowatz, L. a, Jennings, J. D., Lang, J. a, and Kenney, W. L., (2009).
Ketorolac alters blood flow during normothermia but not during hyperthermia in middle-aged human skin.
Journal of applied physiology (Bethesda, Md.: 1985), 107(4), pp.1121–7.
Weblink

Holowatz, L. a, and Kenney, W. L., (2011)a.
Acute localized administration of tetrahydrobiopterin and chronic systemic atorvastatin treatment restore cutaneous microvascular function in hypercholesterolaemic humans.
The Journal of physiology, 589(Pt 19), pp.4787–97.
Weblink

Holowatz, L. a, and Kenney, W. L., (2009).
Chronic low-dose aspirin therapy attenuates reflex cutaneous vasodilation in middle-aged humans.
Journal of applied physiology (Bethesda, Md.: 1985), 106(2), pp.500–5.
Weblink

Holowatz, L. a, and Kenney, W. L., (2011)b.
Oral atorvastatin therapy increases nitric oxide-dependent cutaneous vasodilation in humans by decreasing ascorbate-sensitive oxidants.
American journal of physiology. Regulatory, integrative and comparative physiology, 301(3), pp.R763–8.
Weblink

Holowatz, L. a, and Kenney, W. L., (2007).
Up-regulation of arginase activity contributes to attenuated reflex cutaneous vasodilatation in hypertensive humans.
The Journal of physiology, 581(Pt 2), pp.863–72.

Holowatz, L. a, Santhanam, L., Webb, A., Berkowitz, D. E., and Kenney, W. L., (2011).
Oral atorvastatin therapy restores cutaneous microvascular function by decreasing arginase activity in hypercholesterolaemic humans.
The Journal of physiology, 589(Pt 8), pp.2093–103.

Kuhlenhoelter, A. M., Kim, K., Neff, D., Nie, Y., Blaize, A. N., Wong, B. J., Kuang, S., Stout, J., Song, Q., Gavin, T. P., and Roseguini, B. T., (2016).
Heat therapy promotes the expression of angiogenic regulators in human skeletal muscle.
American Journal of Physiology - Regulatory, Integrative and Comparative Physiology. DOI:10.1152/ajpregu.00134.2016.

Lang, J. a., Holowatz, L. a., and Kenney, W. L., (2009).
Local tetrahydrobiopterin administration augments cutaneous vasoconstriction in aged humans.
The Journal of physiology, 587(Pt 15), pp.3967–74.

Nabavi Nouri, M., Ahmed, A., Bril, V., Orszag, A., Ng, E., Nwe, P., and Perkins, B.a., (2012).
Diabetic neuropathy and axon reflex-mediated neurogenic vasodilatation in type 1 diabetes.
PloS one, 7(4), p.e34807.

Petrofsky, J., Alshammari, F., Khowailed, I. A., Rodrigues, S., Potnis, P., Akerkar, S., Shah, J., Chung, G., and Save, R., (2013).
The effect of acute administration of vitamin D on micro vascular endothelial function in Caucasians and South Asian Indians.
Medical science monitor: international medical journal of experimental and clinical research, 19, pp.641–7.

Wong, B., and Minson, C., (2011).
Altered thermal hyperaemia in human skin by prior desensitization of neurokinin-1 receptors.
Experimental physiology, 96(6), pp.599–609.

Moor Instruments are committed to product development. We reserve the right to change the specifications below without notice.


Analogue outputs

Output parameter: Measured temperature
Range: 0 – 5.0 V
Resolution: 12-bit
Number of outputs: 2
Connectors: 50 Ohm BNC sockets

Dimensions

235x80x200 mm (WxHxD) nominal

Display

Graphical LCD with backlight

Electrical safety classification

Class I
Type BF applied parts

General classification

Protection against ingress of water or particulate matter: IPX0
Not suitable for use in an oxygen rich environment
Not suitable for use with flammable anaesthetics

Indications for use

The moorVMS-HEAT skin heater and temperature monitor is intended for the localised control and temperature measurement of skin. The device is intended for clinical research use. The devices are intended for use by or under the supervision of a physician.

Measured parameters

Temperature (2 Channels)
Range: 5 – 50°C
Accuracy: +0.3°C Precision: 0.1°C
Resolution: 0.1°C ADC resolution: 12-bit
Display update rate: 2 Hz Sampling rate: 40 Hz

Method of sterilisation

Device and probes are not suitable for sterilisation.

Mode of operation

Continuous

Operating environment

Indoor: e.g. hospital or laboratory use Temperature range: 15 – 30°C
Atmospheric pressure: 86 – 106 kPa
Humidity: 0 – 80%, non-condensing
Maximum altitude: 2000 m

Power source

AC mains, 100-230 V, 50-60 Hz, 30 VA

Storage and transportation environment

Temperature range: 5 – 45°C
Atmospheric pressure: 50 – 106k Pa
Humidity: 0 – 80%, non-condensing

Temperature control

Range: 20°C (or ambient) to 45°C
(For HT version – up to 47°C with 10 minute heater cut-out timer when heating above 45°C)
Resolution: 0.1°C
Precision: 0.1°C

Weight

1.6 kg

Moor Instruments生产多种探头,旨在帮助用户测量各种组织的流量。 我们将乐于为您的特定应用提供建议,也希望以下说明能有帮助。