Moor Instruments are committed to product development. We reserve the right to change the specifications below without notice.
Measured parameters: Temperature, 2 channels.
Temperature measurement: Range: 0 – 50 °C
Accuracy 10 – 45 °C : ± 0.3 °C, ± 0.5 °C outside this range ± 0.5 °C over whole range for SHP3 probe
Precision: ± 0.1 °C
Resolution: 0.1 °C
Temperature control: Range: 20°C (or ambient) to 45 °C
Precision: Set temperature ± 0.1 °C within 2 minutes.
Resolution: 0.1 °C (Front panel and serial control). 0.5 °C (Analogue remote control)
Maximum heater power: 5 Watts.
Maximum rate of heating Limited by probe size.
Type of protection against electric shock: Class I Equipment
Degree of protection against electric shock: Type BF Applied part
Medical Devices Directive classification Class IIa (Active device for diagnosis)
Power source: AC Mains power.
Mains supply voltage range: Universal voltage 100 – 230V, 50 – 60Hz
Power rating: 55VA
Degree of protection against flammable anaesthetics: Not Protected.
Degree of protection against ingress of liquid: Probe head of SHP3 probe IPX7 Instrument, all other probes Not protected, IPX0
Mode of operation: Continuous
Operating environmental conditions Temperature: +10°C to +40°C
Relative humidity: 30% to 75% excluding condensation
Atmospheric pressure: 700hPa to 1060hPa
Transportation and storage environmental conditions Temperature: 0°C to +45°C
Relative humidity: 0 to 80% RH Excluding condensation
Atmospheric pressure: 500hPa to 1060hPa