IRLD20 Specifications

Moor Instruments are committed to product development. We reserve the right to change the specifications below without notice.

The IRLD20 infra-red laser diode light source is a Class IIa device under EC directive 93\42EEC 14 June 1993 Medical Device Directive.

CLASSIFICATION

Type of protection against electric shock: Class I equipment
Degree of protection against electric shock: Type BF equipment
Degree of protection against harmful ingress of water: Ordinary equipment (not protected), IPX0
Degree of safety in the presence of a flammable anaesthetic mixture: Not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or with nitrous oxide
Mode of operation: Continuous
Degree of electrical connection between equipment and patient: Equipment with applied part not designed for conductive connection to the patient

OPTICS

Laser Classification: 3B (BS EN 60825-1:1994)
Wavelength: 785nm ± 10nm *
Maximum accessible power: 20mW *
Angular spread of laser light from probe tip: 28° ± 5° *
Nominal ocular hazard distance (MPE for the eye): 50cm

ELECTRICAL

Mains supply: Universal 100 to 230V ~
Power rating: 25 VA

PHYSICAL DIMENSIONS

(W x H x D) mm 250 x 85 x 250 mm
Weight: 2.8 kg

ENVIRONMENTAL STORAGE AND TRANSPORTATION CONDITIONS

Temperature: 0 - 45 °C
Humidity: 0 - 80 %
Non-condensing Atmospheric pressure: 500 - 1060 hPa

OPERATING CONDITIONS

Temperature: 15 - 30 °C
Humidity: 30 - 70%
Non-condensing Atmospheric pressure: 700 - 1060 hPa

* Specifications include cumulative measurement uncertainties and expected increases in values after manufacture.

[Email this page|printer friendly version]