Moor Instruments are committed to product development. We reserve the right to change the specifications below without notice.
OPTICS

Laser Safety Classification
Class 3A per BSEN 60825-1:1994
Class 1M per IEC 60825-1:2001
Class 1 per 21 CFR 1040.10 and 1040.11
Output power 1.6mW max.
Wavelength 785nm + 10nm
Angular spread of laser light from probe tip 30° + 3°
All specifications include cumulative measured uncertainties and expected increases in values after manufacture.
SIGNAL PROCESSING
Bandwidth: high pass 20Hz
Low pass selectable 3KHz, 15KHz, 22KHz
Output time constants: 0.02s, 0.1s, 0.5s, 1.0s, 3.0s
Automatic gain control and zeroing
OUTPUTS
Digital LCD screen display
40Hz maximum data rate
0-1000 arbitrary perfusion units
1 serial interface RS232 optically isolated
8 analogue outputs 0-10v D/A 12 bit (external bus)
1 moorLAB Bus
INTERNAL SOFTWARE FEATURES
Flash memory programmable from PC
Output scale adjustment
Filter bandwidth selection
Calibration - storage of four probe scaling constants per channel
GENERAL
Universal voltage switch-mode power supply range 84 to 264v at 40VA, 50 to 60 Hz
Dimensions W H D mm
SERVER 275 x 60 x 260
SATELLITE 275 x 110 x 280
Weight kgs
SERVER 3.5
SATELLITE 1.5
STORAGE AND TRANSPORT CONDITIONS
Temperature 0-45oC
Humidity 0-80% RH
Atmospheric pressure 500-1060hPa
Type of protection against electric shock - Class 1
Degree of protection against electric shock - Type BF
Degree of protection against ingress of liquid - IPXO (not protected)
Degree of protection against flammable anaesthetics - equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with nitrous oxide.
Mode of operation: continuous.
PERFUSION MEASUREMENTS
FLUX
Accuracy: ± 10%, relative to Moor Instruments ‘standard’ moorLAB
Precision: ± 3% of measurement value
Range: 0-1000 PU
CONC
Accuracy: ± 10%
Precision: ± 5% of measurement value
Range: 0-1000AU
DC
Accuracy: ± 10%
Precision: ± 3%
Range: 0-255